Release Notes 1.1.1.17
Release Date: 21 / 02 / 2024
Firmware: BC43 Version: 2.0.3468 BLE Version: 1.36.22
For version information, please check the prefix at the top left of the screen or refer to About in Preferences
Previously 1.1.1.14 Released: 11 / 12 / 2023
Limitations:
The Width x Height Quick Montage is currently not creating a high speed stitched output.

For this button to create the stitched output please select Montage and Image processing options on the current protocol and turn Apply Stitching On.
See the below selections in purple for the needed settings and the stitched output is in red.

To add the 48 and 384 multi-well template options please reach out to your local Andor Support Representative.
Enhancements
Added support for “load oil” movement with Silicone Oil immersion objectives. Does not yet support Focus Lock for long term live cell imaging.
Added 48 and 384 multi-well template (Available on request)
Reinstated combined view in split views with the ability to replace with a 4th channel.
Added button to toggle a clickable stage view with the joystick when bounds have been set.
Added new buttons to allow users to drive the stage to a position and then set this position as either top left or bottom right bounds.
Bounds can be updated by clicking on a position in the current image and that point being applied to the bounds.
Opened the 3x3 Montage preview button to accept values to a Max Width of 20 and Max Height of 10
Added a button that allows the transfer of top left and bottom right positions on set bounds into protocol perimeter montage.
Contrast level is stored per image.
Easier moving and resizing of small regions.
Taskbar icon reflects protocol status and state now including Deconvolution and Stitching.
Prevent Windows Update and User logout from shutting down Fusion while an acquisition is in progress.
Error message is now displayed if the image file path length exceeds the Windows maximum limit. Protocol will not run until file path name is corrected (shortened).
Service information now stored globally
Bug Fixes
Bounds
Fixed bug where the position of the image was not reflected inside set bounds during live mode.
Protocol
Fixed opening older protocols created in 1.1.0.1 crashes Fusion
Fix added to prevent images not being deleted when stopping a protocol before completion.
Fix for multi position acquisition slow down or hang.
Deconvolution
Fix for Fusion crashes when opening deconvolved image in split-channel MIP.
Fix for deconvolution crashing when trying to load images where the first file in the dataset is missing.
Fix for slow down.
Fix estimated time updates. When deconvolution is activated through the estimated time is now updated.
Fix for deconvolution preview window closing when clicked.
Stitcher
Fix error when commas used as decimal separators.
Stability
Fusion now prevents Windows OS shutdown while deconvolution and/or stitching processing are active.
Fix for reading incorrect Objective Numerical Aperture value.
Fix for crash when editing the imaging mode.
Fixed crash when loading an image file from the root folder.
Fix for null control Unicode character in channel config files.
Improved the text error message displayed when all the fields in a table are deleted.
Fix to prevent invalid configuration files causing unsuccessful initialisation of Fusion
The number of log files retained has been increased to 20.
Fix a crash related to changing between images