Fusion Dragonfly Software Guide

Calibration Information

Viewing Image Calibration Information

The image calibration menu is located in the top right of the image viewing window. Calibration values can be adjusted if required.

Calibration Location.png To view the calibration information menu:

  1. Click the i symbol.

  2. The current information for the image is displayed: Calibration menu.png

  3. Click the pencil icon to open the fields for editing.

  4. The calibration x, y, z start and end points, and voxel size and count values can be modified as required. Calibration edit.png

  5. Click OK or Cancel to complete or cancel editing.


The XY pixel size is determined by the devices in use. For example, the camera, objective information from the microscope, confocal device. If you wish to change or correct the pixel size for all images, try adding the Optical Magnification Component to the Device Hub.

The Z Start and End positions are based on a combination of both Scope (Reference) and Piezo (Scan) positions (if either or both are connected at time of acquisition).